Page 7: Section 1, Functional description, Intr oduction

background image

LC101 CO

2

Module OEM Implementation Man

ual

W

elch Allyn OEM

T

echnologies

C

onfi

dential

P

age

7

Section 1

Functional Description

Intr

oduction

The LC101 Module is designed to control and acquire data from an inter

nal Welch Allyn OEM

Technologies CO

2

bench.

Utilizing a pump to aspirate the patient sample to the CO

2

bench, the

CO

2

concentr

ation in the patient’s expired respiratory gas can be obtained. The sidestream

system includes an internal sidestream CO

2

bench, electronics and pneumatic components

.

The LC101 Module calculates CO

2

measurements and respir

atory rate and outputs this data with

CO

2

w

aveform to the host system via a serial communications interface.

Intended Use

The LC101 Module is intended f

or use as a subsystem within a medical instrument or device.

The Module measures levels of CO

2

contin

uously or intermittently in the sidestream mode. The

Module has been designed to be used on both intubated and non-intubated patients. The Module
and its accessories should always be used as described in the host system’s operator’s manual
and only for the purpose(s) intended. Areas of intended use include:

Hospital Intensive Care Units (ICU)

Hospital Emergency Department/EMS (ED/EMS)

Surgical Operating Rooms (Anesthesia)

Hospital Post Anesthesia Care Units (PACU)

Outpatient Surgical Units

Skilled Nursing Facilities / Subacute Care

Transport (Air, Ground and Sea)

Home Care / Traditional Health Care

The most common user will be a nurse or skilled healthcare professional (such as an EMT,
paramedic, respiratory therapist, clinical engineer, etc.).