Status Message Code Table
Code (hex)
Suggested Message
00
Status OK
01
Invalid Command
02
Invalid Data
03
Unprotected or Unprotected Operation
06
Acknowledge Mode Command
11
CO
2
Sensor Start Up In Progress
15
Vacuum Offset too Large
16
No Watertrap
17
Watertrap Or Cannula Occlusion
18
Exhaust Occlusion Or Pneumatic Leak
21
Calibration Already in Progress
22
Calibration Not in Progress
23
Low Run Time
24
Calibration Ready For Next Step
25
Calibration in Progress
26
Calibration OK
27
Calculation Error
28
Unable to Perform Calibration, Parameters missing
29
Unable to Perform Calibration, Data Error
2A
Bad Calibration CRC
40
Watchdog Error
44
System EEPROM CCITT/CRC Error
46
System FLASH ROM CCITT/CRC Error
47
System Communication Error (LC101 internal)
4B
System External RAM Error
4C
System RAM Error
4D
System FLASH ROM Error/Checksum
4E
Stack Overflow Error
4F
System SW Error/Main Program
51
Manufacturer ID Mismatch
57
Sensor Not Found
60
CO
2
Sensor EEPROM Error/Revision #
65
CO
2
Sensor EEPROM Error/Read/Write
66
CO
2
Sensor EEPROM Error/CCITT/CRC Code
70
CO
2
Sensor Temperature Too High
71
CO
2
Sensor Temperature Too Low
80
Pump Failure
81
Unexpected Reverse Flow
82
Unexpected Forward Flow
84
Barometric Pressure Too High
85
Barometric Pressure Too Low