beautypg.com

Advisories, Lc101 advisories, Pneumatic system advisories – Welch Allyn LC101 CO2 Module - User Manual User Manual

Page 66: Lc101 advisories pneumatic system advisories

background image

LC101 CO

2

Module OEM Implementation Manual

Page 66

Confidential

Welch Allyn OEM Technologies

Advisories

LC101 Advisories

27, Calculation Error

The Module detected an error when calculating the CO

2

measurement. This message is

available in all modes. The error is generated when the calculated CO

2

value is over 99 mmHg.

Pneumatic System Advisories

Status
Message Code
(hex)

Definition

Possible Causes

00

status ok

01

communication error

• host/Module communication is in error

02

host command with invalid data

• identifier is inconsistent with data

03

unprotected operation violation

• host attempts to invoke a protected operation

06

acknowledge command

11

CO

2

sensor startup in progress

• status requested during CO

2

sensor startup

phase

27

calculation error

• calculated CO

2

value is out of range

Status Message Code
(hex)

Definition

Possible Causes

15

vacuum offset too
large

• input connected to an external vacuum source

• faulty receptacle/watertrap receiver switch

16

no watertrap

• watertrap not properly installed

• misaligned watertrap receiver switch

• faulty receptacle/watertrap receiver switch

17

watertrap or cannula
occlusion

• obstructed watertrap/cannula

• sample line kink

• obstructed internal pneumatic component

18

exhaust occlusion or
pneumatic leak

• obstructed exhaust port

• exhaust tubing kink

• faulty/disconnected pump