beautypg.com

Welch Allyn LC101 CO2 Module - User Manual User Manual

Page 23

background image

LC101 CO

2

Module OEM Implementation Manual

Section 3 - LC101 Calibration

Welch Allyn OEM Technologies

Confidential

Page 23

To Perform a Two-Point User Calibration:

The user is required to apply a 0% CO

2

, or “zero” gas, and also apply a gas of known CO

2

concentration (referred to as “span” gas) during the various steps of a Two-Point User
Calibration. The user is required to manually enter the value of the Span Gas applied into the
Monitor.

To perform the Two-Point User Calibration, the user must follow the Original Equipment
Manufacturer’s procedure typically provided in the Operator’s Guide or User Service Manual.
(Refer to illustration on page 22.)

1.

Select a well-ventilated room to perform the calibration.

2.

Let the LC101 run for at least 5 minutes prior to calibration.

3.

For the “Sample Zero Reference Gas” step, attach the CO

2

scrubber to the watertrap inlet

according to the CO

2

scrubber “Directions for Use” (or use a 0% medical-grade gas source).

4.

Proceed with the zero gas phase of the calibration as defined by the host system. Discon-
nect CO

2

scrubber (or 0% gas source) from the watertrap after the zero gas phase and prior

to the next calibration step.

5.

For the “Sample Span Reference Gas” phase of the calibration, provide a medical-grade
gas source with a known CO

2

concentration between 8% and 12%, regulated to a pressure

between 5 and 7 psi.

NOTE Do not apply any pressure directly to the Module inlet or outlet.

6.

Introduce a steady stream of CO

2

gas. Set and monitor the flow rate to approximately

1 liter/minute (±10%). To verify flow, place the bleed line of the calibration sample line in a
glass of water. If bubbles emerge, gas supply is sufficient. If no bubbles emerge, there is an
insufficient supply of gas.

7.

Connect a sample line from the sample elbow, or Tee, as shown above, to the Monitor’s
watertrap inlet. Make sure the bleed line is directed away from the monitor. Allow the refer-
ence gas to flow for at least one minute.

8.

Sample the Span Reference Gas.

9.

After the Two-Point User Calibration steps have been followed, disconnect the LC101 from
the gas supply and test setup.

10. A verification may be performed using the same gas delivery set up. Verify the observed

CO

2

gas reading is within 3 mmHg or 10%, whichever is greater, of the CO

2

value of the test

gas supplied. (If not, refer the user to the appropriate troubleshooting information.)