Configuration
29
Figure 3-7. SCALES Menu, CALIBR Submenu
See the
iQUBE
Installation Manual, PN 77224, for information about configuring
iQUBE
scales.
Table 3-5. SCALES Menu, CALIBR Submenu Parameters
SCALES Menu, CALIBR Submenu
Parameter
Choices
Description
Level 4, CALIBR submenu
WZERO
—
Press ENTER to display and edit the zero calibration A/D count or millivolts value.
WVAL
—
Press ENTER to display and edit the test weight value.
WSPAN
—
Press ENTER to display and edit the span calibration A/D count or millivolts value.
WLIN
POINT 1 —
POINT 5
Press ENTER to display and edit test weight and calibration values for up to five linearization points.
Perform linear calibration only after WZERO and WSPAN have been set.
REZERO
—
Press ENTER to remove an offset value from the zero and span calibrations.
Note
Use this parameter only after WZERO and WSPAN have been set. See Section
48 for more information about using this parameter.
UNITS
lb
kg
g
oz
tn
t
gn
troyoz
troylb
lt
CUSTOM
NONE
OFF
Specifies primary units for displayed and printed weight. Values are: lb=pound; kg=kilogram;
g=gram; oz=ounce; tn=short ton; t=metric ton; gn=grain; troyoz=troy ounce; troylb=troy
pound; lt=long ton.
WZERO
WVAL
WSPAN
WLIN
REZERO
POINT 4
POINT 5
POINT 1
POINT 2
POINT 3
VISIBLE
GRADS FORMAT
CALIBR
…
…
SCALES Menu, FORMAT Submenu, SPLIT
OFF
Parameter
Choices
Description
Table 3-4. SCALES Menu, FORMAT Submenu Parameters, SPLIT
OFF (Continued)