103
10.13 Specifications
Power
Line Voltages
115 or 230 VAC
Frequency
50 or 60 Hz
Power Consumption
25 watts maximum
A/D Specifications
Excitation Voltage
10 ± 0.5 VDC,
16 x 350
or 32 x 700 load cells per A/D card
Sense Amplifier
Differential amplifier with
4- and 6-wire sensing
Analog Signal Input Range –6 mV to +40 mV
Analog Signal Sensitivity
0.3
V/grad minimum @ 7.5 Hz
1.0
V/grad typical @ 120 Hz
4.0
V/grad typical @ 960 Hz
A/D Sample Rate
7.5–960 Hz, software selectable
Input Impedance
>35 M
typical
Internal Resolution
8 000 000 counts
Wt Display Resolution 9,999,999
Input Sensitivity
10 nV per internal count
System Linearity
±0.01% of full scale
Zero Stability
±150 nV/°C, maximum
Span Stability
± 3.5 ppm/°C, maximum
Input Voltage Differential ±800 mV referenced to earth ground
Input Overload
Load cell signal lines ±10 V continuous, ESD protected
RFI/EMI Protection
Communications, signal, excitation, and sense lines protected
Digital Specifications
Microcomputer
Motorola ColdFire
®
MCF5307 main processor @ 90 MHz
Digital I/O
8 I/O channels on CPU board; optional 24-channel I/O expansion card available
Digital Filter
Software selectable: 1–256, enhanced Rattletrap
®
hybrid digital filtering
Serial Communications
Serial Ports
Ports on CPU board support up to 115200 bps; optional dual-channel serial expansion card available
Port 2
Full duplex RS-232 with CTS/RTS
Port 4
Full duplex RS-232; 2-wire RS-485, 20mA