Model 2090 Multi-Device Controller

MODEL 2090 QUICK REFERENCE


Error List

Errors less than

E100

can be cleared by pressing any key on the front panel to acknowledge the error. The

cause of the error condition should be corrected before continuing. The remaining errors may indicate that
the 2090 requires service. If one of these errors occurs, power off the unit for several seconds and power it
back on. If the error recurs, contact EMC Test Systems, L.P. for service.

E001

NVM Initialization Error – Parameters restored to factory defaults due to dead battery.

E002

Motor not moving – Device stalled due to overload, power problems, or failure.

E003

Device will not stop – Device continued moving for too long after motion was disabled.

E004

Device moving wrong direction – Device wired wrong, failing, or reverse delay too short.

E005

Hard Limit hit – Device has reached a mechanical limit.

E006

Polarization limits violation – Tower polarized outside allowed target polarization limits.

E007

Communication lost – No comm. w/ device; power off, fiber disconnected, or failure.

E008

Flotation violation – Attempt to disengage air flotation while turntable is in motion.

E009

Encoder failure – Detected changes in position consistent with an encoder problem.

E010

Trigger failure

E011

Overheat Error – Motor or control board over temperature

E012

Relay Failure – Motor drive control relays not switching properly

E100

ROM test error – ROM has been detected as being corrupt.

E101

RAM test error – RAM has been detected as being corrupt.

E102

NVM test error – NVM has been detected as being corrupt.

E103

Low Power test error – Power test failed.

E104

GPIB test error – GPIB test failed.

E996

BERR Error

E997

Uninitialized Interrupt

E998

Spurious Interrupt

E999

Invalid Interrupt Error

©ETS-Lindgren, April, 2006
Revision G – P#399199