Model 2090 Multi-Device Controller

CONFIGURING FOR USE

Before using the Model 2090, it must be configured to

match the primary devices connected to it. To the Model

2090, all primary devices are generic motor bases, with the

ability to move in two directions, report an encoder

position, and perform a few other basic functions. To

activate behavior specific to a certain type of device, i.e. a

bore sight tower or a two-speed turntable, it is necessary to

enter additional information into the Model 2090’s

configuration parameters table. The previous section

describes the process for entering these parameters. This

section describes each parameter and their typical settings.

The default factory preset configuration is given below.

2090 Default Configuration:

Parameter Value Description

P1 1

Tower

P2 0

Standard (non-bore sight) tower

P3 000

Infinite scan count

P4 00

0 cm polarization offset

P5 0

Standard bore sight correction

P8 0.5

0.5 second reverse delay

P9 8

Primary GPIB address 8

B1 000

User options disabled

c 2000

2000 encoder counts per meter

S0 -1

Step speed = run speed

S1 31

Speed 1 ~12.5% of max speed

S2 63

Speed 2 ~25% of max speed

S3 95

Speed 3 ~37.5% of max speed

S4 127

Speed 4 ~50% of max speed

S5 159

Speed 5 ~62.5% of max speed

S6 191

Speed 6 ~75% of max speed

S7 223

Speed 7 ~87.5% of max speed

S8 255

Speed 8 = max speed

Oc On

Overshoot compensation enabled

ACC 2.0

Acceleration = 2 seconds