Appendix B: Maintenance

Model 2090 Multi-Device Controller

e upgrade file from the program, “

Check

” as it performs

t,

possibly requiring a return to the factory for repair.

Should the upgrade fail for any reason, the motor base

operating firmware will be corrupt. As long as the

interruption occurred prior to the “

SAVE

” cycle, cycling

power on the motor base should restore it to normal

operation. If not, contact ETS-Lindgren for assistance.

Once the upgrade is complete, the motor base will restart,

booting into the new firmware. The Model 2090 will

typically indicate a communication lost (

E007

) error

momentarily until the motor base restores communication.

If the Model 2090 does not automatically recover

Press the “Program” button to initiate the upgrade process.

The flash upgrade file may contain additional

documentation encoded into it, which will be displayed

prior to starting the upgrade process. Once the upgrade is

initiated, the program will display the upgrade status in a

dialog and the Model 2090 will display the status of the

upgrade in the associated device window. The Model 2090

will display “

begin

” as the program negotiates with the

controller and motor base, “

FILE

” to indicate it is receiving

th

some functional checks on the motor base interface board,

prog

” as it sends the firmware data to the motor base,

Verif

” as it verifies the motor base received the file

properly, and “

SAVE

” as it commands the motor base to

store the data to EEPROM and restart. WARNING: Do

not cycle power or otherwise interrupt the firmware

download process, or memory corruption will resul

154

© ETS-Lindgren– April, 2006

Revision G– P# 399199