Model 2090 Multi-Device Controller

For towers equipped with air polarization, pressing the

POLARIZATION key, while the current position is shown

in the display, will toggle the state of the air solenoid valve

in the motor base. The HOR or VERT indicator will light

to indicate the state of the solenoid. If the Polarization

Offset (parameter

P4

on a standard tower) is non-zero, the

value will be added or subtracted from the current position

depending on the target polarization. In moving from

horizontal to vertical polarization, the value will be

subtracted, since the antenna will rotate from being the

offset distance above the centerline of the boom to being

that distance to the side. In moving from vertical to

horizontal, the value will be added to the current position.

The actual motion of the polarization mechanism is

determined by various factors, including available air

pressure and flow control settings on the air polarization

piston cylinder. There is no feedback to indicate that the

cross boom has completed the polarization cycle. The user

should allow a certain amount of time for the polarization

cycle to complete before continuing testing, or use visual

feedback (i.e. from a camera system) to verify that the

polarization operation has completed.

If the current position at the target polarization is outside

the limits of the target polarization, the Model 2090 will

generate a Polarization Limits Violation (

E006

) error and

not allow the tower to polarize. As described in the

configuration information, the Model 2090 provides two

sets of limits, one for each polarization, designed to protect

antenna elements from damage. (The Model 2090 provides