beautypg.com

Nstalling, Lash – ETS-Lindgren 2090 Controller User Manual

Page 151

background image

Model 2090 Multi-Device Controller

download from the ETS-Lindgren web site at

www.ets-

lindgren.com

.

MB-V IDENTIFICATION

The MB-V interface can be identified visually by the

control board enclosure mounted on the outside of the

motorbase with two (2) metal fiber-optic connectors

extending parallel to the main enclosure. From the Model

2090, use the new BASE parameter in the parameter menu

to verify that the motor base has a flash upgradeable MB-

IV board and to determine the existing firmware revision.

WARNING: Under no circumstances should a motor base

with firmware revision less than V1.04 be updated.

Firmware versions earlier than V1.04 indicate an earlier

revision interface board which is incompatible with

firmware revisions V1.04 or later. If you determine you

have an MB-IV motor base with an earlier revision, contact

ETS for more information.

NOTE to 2088 and 5901 users: Upgrading older 2088 or

5901 motor bases will likely cause a change in

performance, slightly increasing the maximum speed. This

increase in speed may also affect positioning accuracy. If

you encounter difficulties after upgrading, contact ETS for

additional information on possible adjustments to your

2088 or 5901 motor base.

©ETS-Lindgren, April 2006

151

Revision G– P#399199