Rs-485 hardware – Cooper Instruments & Systems DFI INFINITY Digital Force Indicator/Controller User Manual
Page 8

Table 2.1. Meter Hookup (RS-232) to the Computer
METER
(DCE)
COMPUTER
(DTE)
PIN SIGNAL/
FUNCTION
RJ-11 RJ-12
D9
D25
RTS, meter from computer
1
2
7
4
TX, meter = RX, computer
2
3
2
3
RX, meter = TX, computer
3
4
3
2
Return 4
5
5
7
NC (not connected)
1,6
(all others)
Table 2.2 shows the pin connection assignments between the RS-232 connection on the meter and the 9-pin or
25-pin “D” connectors of your printer.
Table 2.2. Meter Hookup (RS-232) to the Printer
METER
PIN SIGNAL/
FUNCTION RJ-11
RJ-12
PRINTER
FUNCTIONS
RTS, meter
1
2
Data Terminal Ready (DTR)
TX,
meter
2 3 Received
Data
(RXD)
RX, meter
3
4
Not Connected
Return
4 5 Signal
Return
NC (not connected)
1,6
2.3 RS-485
Hardware
INSTALLATION: The RS-485 card (refer to Figure 2-5A, or if you have an older option card refer to Figure 2-
5B) is the same size and plugs in the same ways as the RS-232 card, refer to Section 2.2.
Figure 2-5A. RS-485 Option Board
JUMPER
RS485
HALF DUPLEX
RS485
FULL DUPLEX
S1-A OPEN
OPEN
S1-B CLOSE
OPEN
S2-A CLOSE
CLOSE
S3-A *
*
CF 125 INFINITY SC GUIDE
4
M1519/N/0605