2mode, 3display mode, Value – Cooper Instruments & Systems DFI INFINITY Digital Force Indicator/Controller User Manual
Page 58: Mode, Display mode

11.2 Mode
Mode is one byte of data (2 characters). It uses the prefixes G, R, and W with the suffix 07. This byte controls
the operating mode of the meter as follows:
0: Batch Mode
1: Rate Mode
2: Sq Rt Mode
The default value is 0, Batch Mode
EXAMPLE: To set the mode to Rate for meter #15, send the command *15W0701
respond with 15W07
15Z05
11.3 Display
Mode
This item is 1 byte (2 characters). It uses command prefixes G and P with the suffix 33, allowing you to read or
change the meter’s display mode.
a) The bit pattern in response to a G33
Table 11.2. Display Mode Bits
FUNCTION
BIT
NO.
VALUE RATE AND SQ RT MODES
BATCH MODE
0-2
0
Not used (Defaults to 1)
Not used (Defaults to 1)
1
RATE is displayed
BATCH is displayed
2
AVG RT is displayed
No. of BATCHes is displayed
3
TOTAL is displayed
TOTAL is displayed
4
TIME is displayed
TIME is displayed
3-6
0
Not used
Not used
1
Not used
Not used
7
0
Not used
RUN mode
1
Not used
Stop mode
(meter is in STOP mode)
b) Bit pattern for a P33 command data format:
FUNCTION
BIT
NO. VALUE
RATE AND SQ RT MODES
BATCH MODE
0-2
0
Not used (Defaults to 1)
Not used (Defaults to 1)
1
Puts RATE on display
Puts BATCH on display
2
Puts AVG RT on display
Puts No. of BATCHes on display
3
Puts TOTAL on display
Puts TOTAL on display
4
Puts TIME on display
Puts TIME on display
5, 6, 7
Not used
Not used
3-7
Not used-Set to zero
Not used
Not used-Set to zero
Not used
EXAMPLE: In RATE MODE, you want to change the display, which is currently showing a rate value, to display
TOTAL on meter #15.
Send command: “15P3303
Meter will send back this message: “15P33
CF 125 INFINITY SC GUIDE
54
M1519/N/0605