3response structure, 1no error, 2echo mode (see sections 10.19 and 11.25) – Cooper Instruments & Systems DFI INFINITY Digital Force Indicator/Controller User Manual
Page 26: 3no echo mode, Response structure, No error, Echo mode (see sections 10.19 and 11.25), No echo mode

2. Suffixes 0E and 0F are not used: the meter will respond to these with an error message.
3. The meter, upon completion of a BLOCK PUT (into RAM) Command, goes to soft reset which does not copy
EEPROM data into RAM.
4. Upon completion of a BLOCK WRITE (into EEPROM) Command, however, the meter goes to hard reset
copying the data from EEPROM into the working RAM. Single PUT (or WRITE) commands to not interrupt the
measurement process of the meter, even when the changes are to scale or offset values.
5. encoding for Write and Put Commands will be described in Section 11 with each Read and Get
response.
8.3 Response
Structure
The meter transmits different response formats according to the type of command it receives, e.g. if it is in echo
or no-echo mode and if an error has occurred or not.
8.3.1 No
Error
8.3.2 Echo Mode (See Sections 10.19 And 11.25)
For type (1) Command:
[nn]ccc[hh]
For type (2) Command:
a) Response for “V01” Command for the Process, Strain Gauge, Temperature, or Universal meters.
[nn][V01][Sa[b]][S
b) Response for “V01” Command for the rate meter/totalizer or batch meters
[nn][V01][Sa][S
c) Response for “X” Command
[nn][ccc]
For type (3) Command:
[nn][ccc]
For type (4) Command:
a) For “P” and “W” Command:
[nn][ccc][hh]
(b) For “Y” Command (Process, Strain Gauge, Temperature and Universal meters only):
[nn][Y01
For type (5) Command:
[nn]ccc[hh]
8.3.3 No Echo Mode
For Commands “P”, “W”, “D”, “E”, “Z”, and “Y”
No response will be transmitted
For type (1) Commands:
[hh]
For type (2) Commands:
a) For “V01” Command:
1) For the Process, Strain Gauge, Temperature and Universal meters:
[Sa[b][S
CF 125 INFINITY SC GUIDE
22
M1519/N/0605