Power
4-3
Always disconnect the battery charger/power adapter from the mains power source
prior to connecting it to or disconnecting it from the battery pack.
The battery charger/power adapter has a built-in LED to indicate the charging status.
Color of LED
Status
Yellow
Battery charging
Green
Battery trickle charging
LED off
Reversed polarity, short circuit, or
no mains connection
Battery status table
Insert the new battery pack into the battery compartment, and insert the plug from the
unit into the connector on the battery pack cable. Replace the lid and secure the lid to
the unit with the screws.
Connecting the unit to the rechargeable battery pack
The maximum capacity of the rechargeable battery will decrease over time, also
reducing the overall battery life. The integrated lifetime calculation function built
into the Portable FlowShark Pulse does not compensate for this reduction.
High and low temperatures and long-term use also lower voltage capacity.
Since rechargeable batteries decrease in capacity over time,
replace the battery pack in the Portable FlowShark Pulse no less
than once every two years. Consider a shorter replacement
period under heavy use.