Battery / rechargeable, Damping – ADS Environmental Services Portable FlowShark Pulse QR 775006 A2 User Manual
Page 137

6-38 ADS Portable FlowShark Pulse Manual
Executing a general reset
Selecting Yes will erase the flash memory.
Saving new values after system reset
Before exiting the menu, the unit seeks confirmation on whether to save new values.
Selecting Yes will reset the Portable FlowShark Pulse to the default parameter
settings.
The default settings will be restored and all customer modifications
will be eliminated (general system reset).
To avoid improper programming and settings, you must execute a
general system reset prior to each initial startup.
Battery / rechargeable
Enter the maximum capacity of the power source that will be used. The system will
use this value when determining the remaining capacity and other information.
Damping
This menu enables you to adjust the display and analog output damping between 5
and 600 seconds.
Damping, Example 1: Damping 30 seconds, jump from 0 l/s to 100 l/s (=100 %) –
unit requires 30 seconds to run from 0 l/s to 100 l/s.