ADS Environmental Services Portable FlowShark Pulse QR 775006 A2 User Manual
Page 175

7-10 ADS Portable FlowShark Pulse Manual
Error
Possible Reason
Potential Corrective Measures
Blank or Flickering
Screen
Connection
Check power supply connection (battery
connector).
Power Supply
Check voltage of the power supply (minimum
of 11.0 volts).
Memory card
Confirm that an ADS-supplied memory card
is in use. Replace card from unauthorized
manufacturer with ADS card, when
applicable.
Verify method used for formatting cards.
Send cards formatted on a PC to ADS.
Sensor Error Indicated
on Display
Connection
Check cable connection.
Battery Voltage
Verify that the voltage is higher than 11.0
volts, and replace the rechargeable battery
pack, if necessary.
DSP Error
Communication
Check communication between CPU or
sensor by pressing the I key. Screen should
display DSP version on the third line.
Erase entire error memory (through the RUN
menu).
When necessary, disconnect the unit from
power for approximately 10 seconds and
restart.
Connection Problems
Contact ADS for verification.
Erratic Readings
Unsuitable Hydraulic
Conditions at
Measurement Location
Check the flow profile on the graph to verify
the quality of the measurement location.
Relocate the sensor to a more suitable
location that exhibits better hydraulic
conditions (e.g., extend calming section).
Remove any existing dirt, sedimentation/silt,
or other obstructions from the front of the
sensor.
Smooth the flow profile by installing
appropriate baffle plates and calming
elements, flow straighteners, or similar
devices upstream from the measurement
location.
Increase damping.
Sensor
Check the sensor orientation (toward the flow
direction and horizontal position).
Inspect the sensor for dirt or blockage.