Parameter Settings
6-63
Measuring countdown Assistant
Once this first measurement is complete, create an obstruction of at least 2.6 inches
(65mm) (ADS recommends 4.7 inches (120 mm)) behind the sensor using a sandbag
or similar object.
The unit will not initiate the second depth measurement at the obstruction until h-
actual exhibits some stability.
Creating obstruction and initiating measurements
Once the readings are stable, the Portable FlowShark Pulse will execute a new 8-
second depth measurement.
Measuring countdown for the second measurement
The unit will display the following readings after the second measurement has been
completed.
h_actual actual depth
h depth before creating an obstruction
v flow velocity reading
Q investigated flow