3 Bedienungslehrgang

IC Metrodata for Win95

56

Select

Equipment setup

of the

Options

menu.

In the

Configured equipment control drivers

window, click on

732 IC

Detector[ COM# ]

and afterwards on

Setup

. The

INITIAL

tab in the

732

IC Detector[ COM# ]

window is opened.

Open the

Manual Control

tab.

Program RUN/STOP

: The program at the 732 IC Detector is

started or stopped.

Valve A FILL/INJECT

: Valve A at the 733 IC Separation Center is

switched to the

Fill

or

Inject

position.

Valve B FILL/INJECT

: Valve B at the 733 IC Separation Center is

switched to the

Fill

or

Inject

position (only

possible with the 733.0020 IC Separation

Center).

Suppressor STEP

:

The suppressor is switched to the next

position (only possible with the 733.0030

IC Separation Center).

Zero ON/OFF

:

The auto-zero is switched on or off.

Pump RUN/STOP

:

The 709 IC Pump is started or stopped.

Log data transmission
The data transmission logging is used for error diagnosis if the com-

munication between PC and 732 IC Detector is faulty. Normally, it

should not be used because the data transmission rate is lowered.

Select

Equipment setup

of the

Options

menu.

In the

Configured equipment control drivers

window, click on

732 IC

Detector[ COM# ]

and afterwards on

Setup

. The

INITIAL

tab in the

732

IC Detector[ COM# ]

window is opened.

Open the

Serial port[ COM# ]

tab.

Select the log type:

Formatted I/O

: The data transmission between PC and 732 IC

Detector 732 is logged line by line.

Byte stream

: The data transmission between PC and 732 IC

Detector 732 is logged character by character

(1 character per line).

File

:

The data transmission logging is saved into the

MemDump.log

file in the program directory.