Metrohm 714 IC Metrodata User Manual
Page 62
3 Bedienungslehrgang
IC Metrodata for Win95
56
•
Select
Equipment setup
of the
Options
menu.
•
In the
Configured equipment control drivers
window, click on
732 IC
Detector[ COM# ]
and afterwards on
Setup
. The
INITIAL
tab in the
732
IC Detector[ COM# ]
window is opened.
•
Open the
Manual Control
tab.
•
Program RUN/STOP
: The program at the 732 IC Detector is
started or stopped.
•
Valve A FILL/INJECT
: Valve A at the 733 IC Separation Center is
switched to the
Fill
or
Inject
position.
•
Valve B FILL/INJECT
: Valve B at the 733 IC Separation Center is
switched to the
Fill
or
Inject
position (only
possible with the 733.0020 IC Separation
Center).
•
Suppressor STEP
:
The suppressor is switched to the next
position (only possible with the 733.0030
IC Separation Center).
•
Zero ON/OFF
:
The auto-zero is switched on or off.
•
Pump RUN/STOP
:
The 709 IC Pump is started or stopped.
Log data transmission
The data transmission logging is used for error diagnosis if the com-
munication between PC and 732 IC Detector is faulty. Normally, it
should not be used because the data transmission rate is lowered.
•
Select
Equipment setup
of the
Options
menu.
•
In the
Configured equipment control drivers
window, click on
732 IC
Detector[ COM# ]
and afterwards on
Setup
. The
INITIAL
tab in the
732
IC Detector[ COM# ]
window is opened.
•
Open the
Serial port[ COM# ]
tab.
•
Select the log type:
Formatted I/O
: The data transmission between PC and 732 IC
Detector 732 is logged line by line.
Byte stream
: The data transmission between PC and 732 IC
Detector 732 is logged character by character
(1 character per line).
File
:
The data transmission logging is saved into the
MemDump.log
file in the program directory.