Metrohm 714 IC Metrodata User Manual
Page 40
3 Learning by doing
IC Metrodata for Win95
34
Modify calibration
•
Select
Open method
of the
File
menu.
•
Select the method used for calibration in the
File Open
window and
click
OK
. An empty chromatogram window appears.
•
Select
Calibration / Graphs
of the
Method
menu. The
Component
win-
dow appears in which the calibration curves and the corresponding
data are displayed.
•
Check with the aid of the graphic representation and the displayed
percentage residual mean standard deviation
RSD
whether the
calibration curve fits optimally the calibration points. Select another
function in the
Formula
field if this is not the case.
•
If you select a new function the message
You attempt to change
GLOBAL parameter! Recalculate all components?
appears.
Click either
Yes
, if all components should be recalculated with this
new function or
No
, if only the displayed component should be re-
calculated.
•
Select one component after the other in the
Component
field and
repeat the procedure to modify the calibration curves.
•
Close the
Component
window by clicking
OK
.