3.9 Manual remote control
IC Metrodata for Win95
55
Write, send and activate a program
•
Select
Equipment setup
of the
Options
menu.
•
In the
Configured equipment control drivers
window, click on
732 IC
Detector[ COM# ]
and afterwards on
Setup
. The
INITIAL
tab in the
732
IC Detector[ COM# ]
window is opened.
•
First open the
Program
tab, then open the
Text
tab. Click the
Add
button to add a new program line.
•
Enter the time
x.x
in min into the first field of the program line.
•
Select the desired program point in the second field (examples see
section 3.8).
•
Select the parameter belonging to the program point in the third
field. For
Flow
,
Pmax
and
Pmin
values must be entered, for
Remote
1
,
0
or
*
must be entered. A certain remote line position can be
changed by moving the cursor in front of this position and then en-
tering
1
,
0
or
*
.
•
Click the
Add
button to add new program lines or click the
Delete
button to delete program lines. Faulty program points are dis-
played in the lower part of the program field.
•
If you wish to use predefined remote instructions, open the
Remote
configuration
tab.
•
Click the
Add
button to add a new remote instruction.
•
Enter any name you like for the remote instruction (max. 19 charac-
ters, e.g.
Ext. Valve Fill
) in the first field.
•
Enter
1
,
0
or
*
for the corresponding remote instruction for remote
control lines 1...8 in the second field. A certain remote line position
can be changed by moving the cursor in front of this position and
then entering
1
,
0
or
*
.
•
Click the
Add
button to add more remote instructions or click the
Delete
button to delete remote instructions.
•
Click the
Activate
button. The entered remote instruction are added
to the standard program points and can also be selected.
•
Open the
Text
tab and add the predefined remote instructions into
the time program.
•
After finishing the programming, click the
Apply
button. The pro-
gram and the parameters are tested for consistency.
•
Switch on the
ENABLED
option.
•
Open the
Initial
tab.
•
Click on
<
. The parameters and the time program are now sent to
the 732 IC Detector.
*
in the lower left part of the window shows
that the PC is sending data. The time program is stored in the 732
IC Detector as program type
cycle
with
number of cycles = 1.
Manual triggering of actions via PC