Metrohm 714 IC Metrodata User Manual
Page 58
3 Learning by doing
IC Metrodata for Win95
52
togram. Therefore, program and parameters can be changed dur-
ing a running method, but the changes are not effective until the
next data acquisition.
Stop a running program
During the program run (green LED in the
Detector is blinking) the
can be stopped as follows:
•
Stop data acquisition, reject chromatogram,
stop and restart program
Click on
or select
Acquire / Suspend chromatogram
to stop data
acquisition.
Click on
or select
Acquire / Restart method
. Confirm the ques-
tion
Available data will be lost. Continue?
with
Yes
. The
Start Run
win-
dow appears. Send the program to the 732 IC Detector by clicking
OK
.
•
Stop data acquisition, save chromatogram, stop program
Close the data acquisition window. Confirm the question
Data aqui-
sition is activ! Standby?
with
Yes
. If the chromatogram should be
saved, confirm the question
01.run not saved. Save?
with
Yes
. Also
confirm the question
Method x modified. Save changes?
with
Yes
.
•
Stop data acquisition, save chromatogram,
continue program
Click on
or select
Acquire / Finish chromatogram
to stop the
data acquisition. The program continues until the
End
flag is set.
If a
Restart method
is set before the program at the 732 IC Detector
has been finished, the following messages appear:
•
Can’t use 732 IC Detector while time program running!
Confirm this
message with
OK
.
•
Detection of hardware failed on
732
IC Detector
. The program at the
732 IC Detector is stopped. This is shown in the upper left part of
the field with
TempStop
. Restart the program with
Acquire / Restart
method
.