3 Learning by doing

IC Metrodata for Win95

46

3.8 Method associated remote control

This section describes the procedure how to associate a time pro-

gram for the 732 IC Detector with a method. In order to do this, a

COM port of the PC and the serial interface of the 732 IC Detector

must have been connected by a 6.2125.060 (25 pin/25 pin) or a

6.2125.110 (25 pin/9 pin) cable. For remote control of the 709 IC

Pump, the pump must have been connected to the 732 IC Detector

by a 6.2125.060 cable (see 732/733 Instruction Manual).

Set RS232 parameters at 732 IC Detector

Set the parameters for the RS232 interface of the 732 IC Detector

as follows:

>CONFIG/RS settings
baud rate: 9600
data bit: 8
stop bit: 1
parity : none
handshake: HWs
RS control: on

Load driver for 732 IC Detector

Select

Equipment setup

of the

Options

menu.

Click on

Load device...

and open the

ic732.equ

file in the

File Open

window.

In the

Select driver for use with 732 IC Detector

window just opened,

double-click the desired

Serial Port

to which the 732 IC Detector is

connected.

Close the

Configured equipment control drivers

window. Confirm the

question

Equipment configuration is modified. Save changes?

with

Yes

.

Associate time program with existing method

Select

Open method

of the

File

menu.

Select the method with whom a time program should be associ-

ated and click

OK

.

Select

Equipment

of the

Method

menu.

Click the

Connect

button in the

Equipment

window.

Double-click

732 IC Detector[ COM# ]

in the

Select driver for use with *

window.

732 IC Detector[ COM# ]

is entered into the

Start with method

field in

the

Equipment

window. This corresponds to a "cycle program" in

the 732 IC Detector. If the program should be started by an "Inject"

at valve A, this row must be moved into the lower

Start with inject

field by clicking the Ð button.