8 method associated remote control – Metrohm 714 IC Metrodata User Manual
Page 52
3 Learning by doing
IC Metrodata for Win95
46
3.8 Method associated remote control
This section describes the procedure how to associate a time pro-
gram for the 732 IC Detector with a method. In order to do this, a
COM port of the PC and the serial interface of the 732 IC Detector
must have been connected by a 6.2125.060 (25 pin/25 pin) or a
6.2125.110 (25 pin/9 pin) cable. For remote control of the 709 IC
Pump, the pump must have been connected to the 732 IC Detector
by a 6.2125.060 cable (see 732/733 Instruction Manual).
Set RS232 parameters at 732 IC Detector
•
Set the parameters for the RS232 interface of the 732 IC Detector
as follows:
>CONFIG/RS settings
baud rate: 9600
data bit: 8
stop bit: 1
parity : none
handshake: HWs
RS control: on
Load driver for 732 IC Detector
•
Select
Equipment setup
of the
Options
menu.
•
Click on
Load device...
and open the
ic732.equ
file in the
File Open
window.
•
In the
Select driver for use with 732 IC Detector
window just opened,
double-click the desired
Serial Port
to which the 732 IC Detector is
connected.
•
Close the
Configured equipment control drivers
window. Confirm the
question
Equipment configuration is modified. Save changes?
with
Yes
.
Associate time program with existing method
•
Select
Open method
of the
File
menu.
•
Select the method with whom a time program should be associ-
ated and click
OK
.
•
Select
Equipment
of the
Method
menu.
•
Click the
Connect
button in the
Equipment
window.
•
Double-click
732 IC Detector[ COM# ]
in the
Select driver for use with *
window.
•
732 IC Detector[ COM# ]
is entered into the
Start with method
field in
the
Equipment
window. This corresponds to a "cycle program" in
the 732 IC Detector. If the program should be started by an "Inject"
at valve A, this row must be moved into the lower
Start with inject
field by clicking the Ð button.