3 Learning by doing

IC Metrodata for Win95

40

Start reprocessing

For reprocessing of the recorded chromatograms proceed accord-

ing to section 3.5.

Start existing queue

Select

Queue/Batch

of the

File

menu.

Select the desired queue

*.que

in the

File open

window and click

OK

.

Select

Clear file names

of the

File

menu to delete all files

*.chw

in the

File

column recorded at the last start of the queue.

Click the

Start

button or select

Run the queue...

to start the proc-

essing of the sample table. The message

Do you really wish to

start the queue?

appears.

Click the

Yes

button. The window for the on-line display of the

chromatogram opens. The status display shows the message

Waiting

. Now the program waits for a start command which is

normally triggered by an Autosampler.

With methods with automatic report output (

Make Report

option in

Method / Method setup / Processing

switched on), pay attention that

the start command by the Autosampler is not triggered before the

report output has been finished.