3.6 Use of sample tables
IC Metrodata for Win95
39
•
Overwrite the fields of the second line with the data for the second
sample.
•
Repeat this procedure until all sample data have been entered.
Start queue
•
Click the
Start
button to start the processing of the sample table.
The window
Do you really wish to start the queue?
appears.
•
Click the
Yes
button. The window for the on-line display of the
chromatogram opens. The status display shows the message
Waiting
. Now the program waits for a start command which is
normally triggered by an Autosampler.
With methods with automatic report output (
Make Report
option in
Method / Method setup / Processing
switched on), pay attention that
the start command by the Autosampler is not triggered before the
report output has been finished.
Open queue during data acquisition
•
Select
Batch
of the
View
menu. The sample data table is displayed,
where all the samples worked off have an entry in the
File
field.
These rows and the first row without file entry (current sample)
cannot be changed during data acquisition.
•
All other rows with samples not worked off can be changed at any
time. For changing the method, select the menu point
File / Change
method
and select the new method from the list of all existing
methods.
•
New rows can be added with
Add Row
, existing rows can be de-
leted with
Del Row
.
•
When you have finished the changes, click the
OK
button. The win-
dow is closed again.