Metrohm 714 IC Metrodata User Manual
Page 17
3.2 Developing a new method
IC Metrodata for Win95
11
•
Set injection valve at 733 IC Separation Center to "FILL" (manually
or by program) and fill sample loop with standard.
•
Set injection valve at 733 IC Separation Center to "INJECT" (manu-
ally or by program). The data acquisition is started automatically.
The status display shows
Measure
, beside it running time, analysis
time and measuring signal (in mV) are displayed. As soon as the
analysis time has passed, the data acquisition is stopped auto-
matically and the chromatogram is integrated.
Modify integration parameters
•
Click on
or select
Integration
of the
Method
menu.
•
As
Delay
, enter the time delay before starting peak integration at
the first peak to be detected.
•
Click the
Apply
button. The chromatogram is reintegrated.
•
Repeat this procedure for all other integration parameters until the
result satisfies your expectations. Click
Apply
after each parameter
changing.
•
Close the
Integration Parameters
window with
OK
.