3.2 Developing a new method

IC Metrodata for Win95

11

Set injection valve at 733 IC Separation Center to "FILL" (manually

or by program) and fill sample loop with standard.

Set injection valve at 733 IC Separation Center to "INJECT" (manu-

ally or by program). The data acquisition is started automatically.

The status display shows

Measure

, beside it running time, analysis

time and measuring signal (in mV) are displayed. As soon as the

analysis time has passed, the data acquisition is stopped auto-

matically and the chromatogram is integrated.

Modify integration parameters

Click on

or select

Integration

of the

Method

menu.

As

Delay

, enter the time delay before starting peak integration at

the first peak to be detected.

Click the

Apply

button. The chromatogram is reintegrated.

Repeat this procedure for all other integration parameters until the

result satisfies your expectations. Click

Apply

after each parameter

changing.

Close the

Integration Parameters

window with

OK

.