3.4 Multipoint calibration
IC Metrodata for Win95
29
3.4 Multipoint
calibration
This section describes the procedure how to perform a calibration
with multiple standards using either an existing or an user-developed
method and to save this multipoint calibration as a new method.
Preparations
•
Insert separating column in IC system and condition it.
•
Select the full scale range at the 732 IC Detector so that all peaks
of the standard with the highest concentration will be inside the
measuring range.
•
Prepare standards and sample.
Record first standard chromatogram
•
If you wish to develop a new method based on a standard method
proceed according to section 3.1 (without sample recording).
•
If you wish to use an existing example method proceed according
to section 3.2 (without sample recording).
•
In both cases, you receive the new method
*.mtw
and the first
standard chromatogram
*.chw
.
Record second standard chromatogram
•
Click on
or select
Load method and run
from the
Acquire
menu.
•
Select the method saved on recording the first standard chroma-
togram in the
File Open
window and click the
OK
button.
•
In the
Start Run
window, enter a name as
Ident
to characterize the
standard (max. 14 characters, e.g. "standard #").
•
Click the
OK
button.
The window for the on-line display of the chromatogram opens.
The status display shows the message
Waiting
.
•
Set injection valve at 733 IC Separation Center to "FILL" (manually
or by program) and fill sample loop with standard.
•
Set injection valve at 733 IC Separation Center to "INJECT" (manu-
ally or by program). The data acquisition is started automatically
and stopped as soon as the analysis time has passed. The chro-
matogram is saved and deleted in the chromatogram window. The
status display shows the message
Waiting
again.