3.4 Multipoint calibration

IC Metrodata for Win95

29

3.4 Multipoint

calibration

This section describes the procedure how to perform a calibration

with multiple standards using either an existing or an user-developed

method and to save this multipoint calibration as a new method.

Preparations

Insert separating column in IC system and condition it.

Select the full scale range at the 732 IC Detector so that all peaks

of the standard with the highest concentration will be inside the

measuring range.

Prepare standards and sample.

Record first standard chromatogram

If you wish to develop a new method based on a standard method

proceed according to section 3.1 (without sample recording).

If you wish to use an existing example method proceed according

to section 3.2 (without sample recording).

In both cases, you receive the new method

*.mtw

and the first

standard chromatogram

*.chw

.

Record second standard chromatogram

Click on

or select

Load method and run

from the

Acquire

menu.

Select the method saved on recording the first standard chroma-

togram in the

File Open

window and click the

OK

button.

In the

Start Run

window, enter a name as

Ident

to characterize the

standard (max. 14 characters, e.g. "standard #").

Click the

OK

button.

The window for the on-line display of the chromatogram opens.

The status display shows the message

Waiting

.

Set injection valve at 733 IC Separation Center to "FILL" (manually

or by program) and fill sample loop with standard.

Set injection valve at 733 IC Separation Center to "INJECT" (manu-

ally or by program). The data acquisition is started automatically

and stopped as soon as the analysis time has passed. The chro-

matogram is saved and deleted in the chromatogram window. The

status display shows the message

Waiting

again.