Page 16

3 Learning by doing
IC Metrodata for Win95
10
3.2 Developing a new method
This section describes the procedure how to develop a new, user-
specific method based on a standard method.
Preparation
•
Insert separating column in IC system and condition it.
•
Prepare standards and sample.
Load standard method and run
•
Click on
or select
Load method and run
from the
Acquire
menu.
•
Select the file
1m732b1.mtw
in the
File Open
window and click the
OK
button.
•
In the
Start Run
window, enter a sample name as
Ident
to character-
ize the chromatogram (max. 14 characters, e.g. "standard").
•
Enter the estimated analysis time as
Duration
.
•
Select the tabs
Sample
,
Eluent
,
Measure
and
Processing
to enter
more parameters to characterize sample and acquirement condi-
tions if required.
•
Click the
OK
button.
The window for the on-line display of the chromatogram opens.
The window title line contains sample name, method and run num-
ber (e.g.
standard (1m732b1.mtw)01.run
). The status display below
shows the message
Waiting
.