Metrohm 714 IC Metrodata User Manual
Page 56
3 Learning by doing
IC Metrodata for Win95
50
Start cycle program and data acquisition
•
Click on
or select
Restart method
of the
Acquire
menu.
•
Update the parameters in the
Start Run
window and start method
with
OK
.
•
The parameters stored in the method and the time program are
sent to the 732 IC Detector.
*
in the lower left part of the window
shows that the PC is sending data.
•
As soon as the program at the 732 IC Detector is ready, it will be
started automatically. This is indicated by the blinking LED in the
figuration in the
Measure
tab, the data acquisition proceeds as fol-
lows:
Start mode = Extern
→
Data acquisition is started with the
Inject
instruction in the time program.
Start mode = Manual
→
Data acquisition is started as soon as
the time program at the 732 IC Detector is running.
•
If the
Restart Method
option is selected in the
Method / Method Setup /
Processing
menu, the time program is restarted at the end of the
program. If the
Restart Method
option is not selected in the
Method /
Method Setup / Processing
menu, only one cycle of this program is
performed. If the method is started from a queue, the program is
stopped as soon as all rows of method X are worked off.
•
The time program is stored as program type
cycle
with
number of
cycles = 1
. The time program and the parameters are sent again to
the 732 IC Detector at the end of the chromatogram. Therefore,
program and parameters can be changed during a running
method, but the changes are not effective until the next cycle.