Page 56

background image

3 Learning by doing

IC Metrodata for Win95

50

Start cycle program and data acquisition

Click on

or select

Restart method

of the

Acquire

menu.

Update the parameters in the

Start Run

window and start method

with

OK

.

The parameters stored in the method and the time program are

sent to the 732 IC Detector.

*

in the lower left part of the window

shows that the PC is sending data.

As soon as the program at the 732 IC Detector is ready, it will be

started automatically. This is indicated by the blinking LED in the

key at the 732 IC Detector. Depending on the con-

figuration in the

Measure

tab, the data acquisition proceeds as fol-

lows:

Start mode = Extern

Data acquisition is started with the

Inject

instruction in the time program.

Start mode = Manual

Data acquisition is started as soon as

the time program at the 732 IC Detector is running.

If the

Restart Method

option is selected in the

Method / Method Setup /

Processing

menu, the time program is restarted at the end of the

program. If the

Restart Method

option is not selected in the

Method /

Method Setup / Processing

menu, only one cycle of this program is

performed. If the method is started from a queue, the program is

stopped as soon as all rows of method X are worked off.

The time program is stored as program type

cycle

with

number of

cycles = 1

. The time program and the parameters are sent again to

the 732 IC Detector at the end of the chromatogram. Therefore,

program and parameters can be changed during a running

method, but the changes are not effective until the next cycle.