3 Learning by doing
IC Metrodata for Win95
48
•
Select the IC system type in the
Initial
tab. With
Anion
and
Cation
temperature coefficient and polarity are set automatically, with
User
these values can be combined independently.
•
Change the parameters for 732 IC Detector, 733 IC Separation
Center and 709 IC Pump to the values to be set for these instru-
ments at the start of the method.
•
First open the
Program
tab, then open the
Text
tab. Click the
Add
button to add a new program line.
•
Enter the time
x.x
in min into the first field of the program line.
•
Select the desired program point in the second field.
•
Select the parameter belonging to the program point in the third
field. For
Flow
,
Pmax
and
Pmin
values must be entered, for
Remote
1
,
0
or
*
must be entered. A certain remote line position can be
changed by moving the cursor in front of this position and then en-
tering
1
,
0
or
*
.
•
Click the
Add
button to add new program lines or click the
Delete
button to delete program lines. Faulty program points are dis-
played in the lower part of the program field.
•
If you wish to use predefined remote instructions, open the
Remote
configuration
tab.
•
Click the
Add
button to add a new remote instruction.
•
Enter any name you like for the remote instruction (max. 19 charac-
ters, e.g.
Ext. Valve Fill
) in the first field.
•
Enter
1
,
0
or
*
for the corresponding remote instruction for remote
control lines 1...8 in the second field. A certain remote line position
can be changed by moving the cursor in front of this position and
then entering
1
,
0
or
*
.
•
Click the
Add
button to add more remote instructions or click the
Delete
button to delete remote instructions.
•
Click the
Activate
button. The entered remote instruction are added
to the standard program points and can also be selected.
•
Open the
Text
tab and add the predefined remote instructions into
the time program.
•
After finishing the programming, click the
Apply
button. The pro-
gram and the parameters are tested for consistency.
•
Switch on the
ENABLED
option and close the window with
OK
.
•
Close the
Equipment
window with
Close
.
•
Select
Save method as
of the
File
menu and save the changed
method under the same or a new name.