3.4 Multipoint calibration

IC Metrodata for Win95

35

Save method

Select

Save method as

of the

File

menu. The

File Save As

window

appears.

Click the

OK

button. A window appears with the message

c:\..... al-

ready exists. Overwrite?

Click the

OK

button.

Record a sample chromatogram

Click on

or select

Restart method

of the

Acquire

menu.

Enter a sample name (max. 14 characters, e.g. "drinking water") as

Ident

in the

Start Run

window.

If required, enter more parameters (e.g.

Dilution

) to characterize

sample and acquirement conditions.

Click the

OK

button.

The window for the on-line display of the chromatogram opens.

The status display below shows the message

Waiting

.

Set the injection valve at the 733 IC Separation Center to "FILL"

(manually or by program) and fill the sample loop with sample.

Set the injection valve at the 733 IC Separation Center to "INJECT"

(manually or by program). The data acquisition is started automati-

cally and stopped as the analysis time has passed.