3.9 Manual remote control
IC Metrodata for Win95
53
3.9 Manual remote control
This section describes the procedure how to send parameters and
time programs to the 732 IC Detector and trigger manually certain
functions. In order to do this, a COM port of the PC and the serial in-
terface of the 732 IC Detector must have been connected by a
6.2125.060 (25 pin/25 pin) or a 6.2125.110 (25 pin/9 pin) cable. For
remote control of the 709 IC Pump, the pump must have been con-
nected to the 732 IC Detector by a 6.2125.060 cable (see 732/733 In-
struction Manual).
Set RS232 parameters at 732 IC Detector
•
Set the parameters for the RS232 interface of the 732 IC Detector
as follows:
>CONFIG/RS settings
baud rate: 9600
data bit: 8
stop bit: 1
parity : none
handshake: HWs
RS control: on
Load driver for 732 IC Detector
•
Select
Equipment setup
of the
Options
menu.
•
Click on
Load device...
and open the
ic732.equ
file in the
File Open
window.
•
In the
Select driver for use with 732 IC Detector
window just opened,
double-click the desired
Serial Port
to which the 732 IC Detector is
connected.
•
Close the
Configured equipment control drivers
window. Confirm the
question
Equipment configuration is modified. Save changes?
with
Yes
.
Send and receive parameters
•
Select
Equipment setup
of the
Options
menu.
•
Click on
732 IC Detector[ COM# ]
in the
Configured equipment control
drivers
window. Click the
Setup
button. The
INITIAL
tab in the
732 IC
Detector[ COM# ]
window is opened.
•
Click on
Ask
. The current parameters of the 732 IC Detector are en-
tered into the
Actual value
fields.
*
in the lower left part of the win-
dow shows that the PC is receiving data.
•
Click on
>
. The parameters are now entered into the fields at the
right side of the
Initial
tab.
•
Select the IC system type in the
Initial
tab. With
Anion
and
Cation
temperature coefficient and polarity are set automatically, with
User
these values can be combined independently.