Metrohm 714 IC Metrodata User Manual
Page 29
3.3 Use of existing methods
IC Metrodata for Win95
23
Preparations
•
Insert separating column in IC system and condition it.
•
Prepare standards and sample.
Load standard method and run
•
Click on
or select
Load method and run
from the
Acquire
menu.
•
Select the desired method (e.g.
dual1sup.mtw
) in the
File Open
win-
dow and click the
OK
button. The window
Start Run
appears.
•
Enter a name as
Ident
to characterize the standard (max. 14 char-
acters, e.g. "standard").
•
Change the analysis time
Duration
if required.
•
Enter more parameters to characterize sample and acquirement
conditions if required.
•
Click the
OK
button.
The window for the on-line display of the chromatogram opens.
The window title line contains sample name, method and run num-
ber (e.g.
standard (dual1sup.mtw)01.run
). The status display below
shows the message
Waiting
.
•
Set injection valve at 733 IC Separation Center to "FILL" (manually
or by program) and fill sample loop with standard.
•
Set injection valve at 733 IC Separation Center to "INJECT" (manu-
ally or by program). The data acquisition is started automatically
and stopped as soon as the analysis time has passed. As the
Re-
start method
option is switched on in the method, the chroma-
togram is saved automatically and deleted in the chromatogram
window.