Metrohm 714 IC Metrodata User Manual
Page 21

3.2 Developing a new method
IC Metrodata for Win95
15
Set processing options
•
Click on
or select
Method setup
of the
Method
menu. Select the
Processing
tab in the
Method Setup
window.
•
The options
Switch to chromatography
(return to «IC Metrodata for
Win95» after finishing of the chromatogram if the program runs in
background),
Save chromatogram
(automatic saving) and
Automatic
Show All
(automatic scaling of axes) are selected by default.
•
If you click the
Make Report
option, the report defined under
Report
Options
will be printed out automatically after finishing of the chro-
matogram.
•
Click the
Restart method
option. This option opens automatically a
window for acquiring a new chromatogram after finishing the old
one.
•
If you click the option
Close window
in addition, the data acquisition
window is closed automatically after finishing of the method.
•
In the
Chromatograms directory
window, a directory path for saving
the chromatogram can be entered.
•
Close the window by clicking
OK
.
Save new method
•
Select
Save method as
of the
File
menu.
•
Enter a name
*.mtw
for the new method in the
File Save As
window
and click the
OK
button.
The
1m732b1.mtw
method is not write-protected!