6 use of sample tables – Metrohm 714 IC Metrodata User Manual
Page 44
3 Learning by doing
IC Metrodata for Win95
38
3.6 Use of sample tables
This section describes how to construct sample data tables with the
aid of a queue. These tables can be processed automatically and are
suited for operation with an Autosampler.
Create new queue
•
Select
Queue/Batch
of the
File
menu.
•
Enter a name for the new queue in the
Filename
field and click
OK
.
•
In the
File open
window, select the method to be used in the queue.
A window with a sample table appears, which contains a row for
sample data entry.
•
Enter a sample name in the
Ident
field.
•
Enter the vial number on the Autosampler in the
Vial
field.
•
Enter the volume of the sample loop in the
Volume
field (only if the
volume is changed).
•
Enter the dilution factor for the sample in the
Dilution
field.
•
Enter the folowing numbers in the
Level
field:
0 for
samples
≥
1
for standards (the level concentrations have
to be defined in advance in the method)
•
Enter the name
*.mtw
of the method for processing the sample in
the
Method
field (the method name has to be defined in advance).
•
The
File
field needs no input (the name of the saved file is entered
automatically in this field at the end of the data acquisition).
•
When all data are filled in in the first line, click the
Add Row
button
to insert a second line for entering sample data. At the same mo-
ment the data of the first line are copied into the second line.