3 Learning by doing
IC Metrodata for Win95
30
Record third to x
th
standard chromatogram
•
Click on
or select
Passport
of the
Method
menu.
•
In the
General
tab, enter a new name as
Ident
characterizing the
standard (e.g. "standard #") and click
OK
.
•
Set injection valve at 733 IC Separation Center to "FILL" (manually
or by program) and fill sample loop with standard.
•
Set injection valve at 733 IC Separation Center to "INJECT" (manu-
ally or by program). The data acquisition is started automatically
and stopped as soon as the analysis time has passed. The chro-
matogram is saved and deleted in the chromatogram window. The
status display shows the message
Waiting
again.
•
Repeat this procedure until the last standard has been recorded.
•
Close the empty chromatogram window. The message
Data acqui-
sition is active! Standby?
appears.
•
Click the
Yes
button.