3.5 Modify several chromatograms

IC Metrodata for Win95

37

Select the desired method for recalculation of all chromatograms

recorded with this method in the

Chromatograms acquired by method

field.

If the sample chromatograms (Level = 0) should be recalculated

switch on the

Sample runs

option.

If the standard chromatograms (Level

1) should be recalculated

switch on the

Calibration runs

option.

If the chromatograms should be reintegrated switch on the

Reinte-

grate

option (if the queue contains standards, the

Recalibrate

option

is switched on automatically). When the reprocessing is started the

Integration parameters

window appears (procedure see section 3.2

– Modify integration parameters).

If a new calibration should be performed switch on the

Recalibrate

option. In this case, select the desired options in the

Recalibration

part of the window.

If a report should be printed out automatically for all recalculated

chromatograms switch on the

Print report

option. When the reproc-

essing is started the

Report options

window appears (procedure

see section 3.2 – Change report options).

Click the

Start

button.