Metrohm 714 IC Metrodata User Manual
Page 43

3.5 Modify several chromatograms
IC Metrodata for Win95
37
•
Select the desired method for recalculation of all chromatograms
recorded with this method in the
Chromatograms acquired by method
field.
•
If the sample chromatograms (Level = 0) should be recalculated
switch on the
Sample runs
option.
•
If the standard chromatograms (Level
≥
1) should be recalculated
switch on the
Calibration runs
option.
•
If the chromatograms should be reintegrated switch on the
Reinte-
grate
option (if the queue contains standards, the
Recalibrate
option
is switched on automatically). When the reprocessing is started the
Integration parameters
window appears (procedure see section 3.2
– Modify integration parameters).
•
If a new calibration should be performed switch on the
Recalibrate
option. In this case, select the desired options in the
Recalibration
part of the window.
•
If a report should be printed out automatically for all recalculated
chromatograms switch on the
Print report
option. When the reproc-
essing is started the
Report options
window appears (procedure
see section 3.2 – Change report options).
•
Click the
Start
button.