3.4 Multipoint calibration
IC Metrodata for Win95
31
Enter concentrations for calibration
•
Select
Open method
of the
File
menu.
•
Select the previously saved method in the
File Open
window and
click
OK
.
•
Select
Calibration / Concentrations
of the
Method
menu.
•
Click the
Add
button and confirm the suggested number
2
for the
Level
in the
Add level
window by clicking
OK
.
•
Enter the concentrations of all components of the second standard
in the new
Level 2
column.
•
Repeat this procedure for every further standard.
•
If all standards have been entered, close the
Concentrations
win-
dow by clicking
OK
.
•
Close the empty chromatogram window. The message
Method
*.mtw was modified. Save changes?
appears.
•
Click the
Yes
button. The
File Save As
window appears.
•
Click the
OK
button. A window appears with the message
c:\..... al-
ready exists. Overwrite?
•
Click the
OK
button.