5.4 Command reference
774 Oven Sample Processor, Instructions for use
83
DEF
0
DEF
Redefining specific instrument settings
During method processing the most diverse settings can be made using
the following DEF commands. The individual entries are accessed by
repeatedly pressing the DEF key (roll-up selection).
>sample sequence
14 DOSRATE
1
160
1…12
0.01…160 ml/min
Dosing speed
The dosing speed for all 12 dosing units can be individually set. The first
parameter stands for the number of the dosing unit and the second pa-
rameter allows setting the dosing speed in ml/min.
>sample sequence
12 FILLRATE 1
160
1…12
0.01…160 ml/min
Filling speed
The filling speed can be set for all 12 dosing units individually. The first pa-
rameter stands for the number of the dosing unit and the second parameter
allows setting the filling speed in ml/min.
>sample sequence
16 LIFTRATE
1 12 mm/s
1 3…12 mm/s
Lift rate
The first parameter of this command defines the lift
on which the command is to be executed. Since
the 774 Oven Sample Processor is equipped only
with Lift 1, it is not practical to enter any other
value at this point.
>sample sequence
17 SHIFTRATE: auto. 20
auto.,+,– 3…20 w/s
Turning direction and speed
Turning direction and speed of the sample rack can be changed as de-
sired. The first parameter determines the turning direction.
auto. :
The changer automatically determines the shortest path.
+ :
The sample rack turns in the counterclockwise direction (to a
higher rack position)
- :
The sample rack turns in the clockwise direction (to a lower
rack position)
The second parameter determines the turning speed in angular de-
grees/sec.