5.10 The remote control tree
774 Oven Sample Processor, Instructions for use
109
:
├
.WetPart
Dosing unit definitions
│ ├
.WetPartNo
Dosing unit no.
1...12
│
∗
│ ├
.1
Port number 1
-
│ │ ├
.MaxRate
Max. dosing rate
0.01...160 ml/min
│ │ ├
.Length
Tubing length
0...1000...30000 mm
│ │ └
.Diameter
Tubing diameter
0...2...20 mm
│ │
│ ├
.2
Port number 2
-
│ │ ├
.MaxRate
Max. dosing rate
0.01…160 ml/min
│ │ ├
.Length
Tubing length
0...250...30000 mm
│ │ └
.Diameter
Tubing diameter
0.1…2…20 mm
│ │
│ ├
.3
Port number 3
-
│ │ ├
.MaxRate
Max. dosing rate
0.01...160 ml/min
│ │ ├
.Length
Tubing length
0...1000...30000 mm
│ │ └
.Diameter
Tubing diameter
0.1...2...20 mm
│ │
│ └
.4
Port number 4
│ ├
.MaxRate
Max. dosing rate
0.01...160 ml/min
│ ├
.Length
Tubing length
0...1000...30000 mm
│ └
.Diameter
Tubing diameter
0.1...2...20 mm
│
└
.Rsset
RS232 Settings
├
.Baud
Baud rate
300, 600, 1200,
│
2400, 4800, 9600
├
.DataBit
Number of data bits
7, 8
├
.StopBit
Number of stop bits
1, 2
├
.Parity Parity
even,
odd,
none
├
.Handsh Handshake
Hws, HWfull,
│
SWchar, SWline,
│
none
└
.CharSet Character
set
IBM, HP, Epson,
Seiko,
Citizen
5.10.4 &Info
Object Description
Input
range
Reference
& Root
:
├
Info
Current data
-
:
├
.Report
Report definition
│ └
.Select Report
type
short, full, config, param,
│
usermeth,
all
│
└
.DetermData
Determination data
├
.Mplist
Meas. point list of temperature
-
:
├
.1
Meas. point 1
-
:
└
.Y
Temperature in °C
read only
:
└
.400
Meas. point 400