Metrohm 774 Oven Sample Processor User Manual
Page 76

5 Detailed description
774 Oven Sample Processor, Instructions for use
70
parameters
>report
Submenu for report type definition
Open the submenu with
>report
report: none
Setting of the report type
none
fulll
short
config
param
usermeth
all
= No report output
= Output of the oven temperatures
= Short result report
= Listing of the device configuration
= Method listing
= List of all stored methods
= Printout of all above reports
When selecting a result report (full or short), a separate report is printed out
for each sequence (start, sample and final sequence).
>report
rec. interval
10 s
Time interval for recording the oven
temperatures
1…10…60 s
Maximum 400 temperature values (meas-
urement points) can be recorded per se-
quence. This results in a maximum re-
cording time of approx. 6…7 hours in the
case of one measuring report per minute.
The recording starts at the start of a se-
quence in each case.
parameters
>changer settings
Submenu for the changer settings
Open the submenu with
>changer settings
rack number
0
The rack that is assigned to the
method
0…16
0 = no particular rack
This setting can force the use of a certain rack with the method chosen. If this
is not desired, the rack number 0 must be chosen.
>changer settings
lift rate 1
12 mm/s
Stroke speed of lift 1
3…12 mm/s
>changer settings
lift rate 2
12 mm/s
Stroke speed of lift 2
3…12 mm/s
>changer settings
shift rate.
20
Turning speed of the rack in angular
degrees/second
3…20
>changer settings
shift direction:
auto.
Turning direction of the rack
+, -, auto.
auto. =
the sample changer chooses
the shortest path for turning.