5 Detailed description
774 Oven Sample Processor, Instructions for use
130
5.11.7 &Diagnosis …
5.11.7.1
Diagnose.Init $G
Diagnose.Init.Select
param, config,
setup, assembly, all
RAM initialization. Sets all default values for the selected sub-branch. See
also 5.11.4.11.
&Diagnose.Init;$G triggers the initialization..
5.11.7.2
Diagnose.RamTest $G
Diagnose.LcdTest
$G, $S, $H
Diagnose.ContrastTest $G,
$S
Diagnose.KeyTest $G,
$S
Diagnose.IoTest $G,
$S
Diagnose.RsTest $G,
$S
Diagnose.EbusTest $G,
$S
Diagnose.BeeperTest $G,
$S
Diagnose.RackcodeTest $G,
$S
Diagnose.FunctionTest
Diagnostic functions. These functions can be triggered with $G. They may
be stopped with $S.
.RamTest
Working memory test
.LcdTest Display
test
.ContrastTest
Display contrast test
.KeyTest
Displays key code and function of each key
.IoTest
Remote interface test
.RsTest
(This function is not applicable via RS232 interface)
.EbusTest
External Bus interface test
.BeeperTest
Test of the acoustic signal (only with
Config.Beeper"on", see 4.10.2.14)
.RackcodeTest Test of rack code recognition
.FunctionTest
Metrohm internal test
5.11.7.3
Diagnose.SimulateKey
0, 1…6, 8…31
Key simulation (Key codes see 5.11.4.2).
5.11.7.4
Diagnose.InstrNo
Instrument number. This entry is not accessible via RS232 interface.
5.11.7.5
Diagnose.OvenTest $G,
$S
Diagnose.Adjust $G,
$S
Diagnose.MeasTest $G,
$S
.OvenTest Heater
test.
.Adjust
Adjustment of temperature and gas flow measurement.
.MeasTest
Test of temperature measurement.
5.11.7.6
Diagnose.PowerOn $G
Simulation of 'Power on'.