2 &config – Metrohm 774 Oven Sample Processor User Manual
Page 125

5.11 Description of the remote control commands
774 Oven Sample Processor, Instructions for use
119
5.11.2 &Config …
5.11.2.1
Config.Aux.Language
english, deutsch, français, español
Config.Aux.Contrast 0…3…7
Config.Aux.Beeper
on, off
Config.Aux.DevName
8 ASCII characters
Config.Aux.Prog read
only
Config.Aux.MaxLift
0...90...100 mm
Config.Aux.MonBeak on, off
Basic configuration / miscellaneous
Language: Selection
of dialog language
Contrast:
Display contrast in steps from 0 to 15
Beeper:
Acoustic warning signal on/off
DevName:
Instrument name as identification for connections with
other instruments (8 ASCII characters, no special charac-
ters)
Prog:
Program version ($Q sends "730.0013")
MaxLift:
Max. Lift height i.e. lowest possible lift position
∗
MonBeak:
Beaker sensor on/off ∗
∗
Modifications effect after RESET or power on.
5.11.2.2
Config.OvenSet.InitTemp
Off, 50…250
Config.OvenSet.MaxTemp 50…275
Config.OvenSet.TempCorr -10…0…10
The initial temperature of the oven control is set on power-up of the unit.
When the maximum temperature is reached, this aborts the running method
and the RS error message E165.
The temperature correction varies the preset temperature for the oven con-
trol, i.e. it also influences the control algorithm.
5.11.2.3
Config.RackDef.RackNo
1...16
Rack number
5.11.2.4
Config.RackDef.Code
000001b...111111b
Config.RackDef.Type
8 ASCII characters
Config.RackDef.WorkH
0...100 mm
Config.RackDef.RinseH
0...100 mm
Config.RackDef.ShiftH
0...20 mm
Config.RackDef.SpecialH
0...100 mm
Rack definitions. Depending on the current rack number (see 5.11.2.3) the
corresponding one of the 16 possible data sets is overlayed.
Code:
Identification code of the rack, has to be unique.
Type:
Rack type. Valid entries are the names of the position ta-
bles.
WorkH:
Working position of the lift in mm from the upper stop of
the tower.
RinseH:
Rinsing position of the lift in mm from the upper stop of the
tower.
ShiftH:
Shifting position of the lift in mm from the upper stop of the
tower.
SpecialH:
Special position of the lift in mm from the upper stop of the
tower.