5.12 Properties of the RS232 Interface

774 Oven Sample Processor, Instructions for use

133

774 Oven Sample Processor as Sender:

Time

RxD

TxD

Data output

Data input

LF

LF

LF

Data output
disabled

Data output
enabled

Request

1.Line

Response

XON

XOFF

external device

Sample Changer

Changer transmission can be stopped by external instruments with XOFF.
After XOFF is received the changer completes sending the line already
started. If data output is disabled for more than 3 s by XOFF, E43 appears
in the display.


Hardware-Handshake, HWs

774 Oven Sample Processor as Receiver:

Time

DTR

LF

RxD

RxD

external device

Sample Changer

DTR


774 Oven Sample Processor as Sender:

Time

RTS

RTS

LF

TxD

TxD

CTS

CTS

external device

Sample Changer

The data flow can be interrupted by deactivating the CTS line.