Metrohm 774 Oven Sample Processor User Manual
Page 72

5 Detailed description
774 Oven Sample Processor, Instructions for use
66
>dosing units
1
max. rate 1
160 ml/min
Max. dosing speed (depends on bu-
rette size)
0.01…160 ml/min
>dosing units
1
tube length 1
1000 mm
Length of the tubing on Dosino port 1
0,
1000, 30000 mm
>dosing units
1
tube diameter 1 2 mm
Diameter of the tubing on Dosino port
1
0.1…2.0…20 mm
>dosing units
1
max. rate 2
160 ml/min
Max. dosing speed (dependends on
burette size)
0.01…160 ml/min
>dosing units
1
...........................until Port 4
Enter the tubing diameter for all four
ports of a Dosino
Only the dosing speed is relevant for the 685 Dosimat; the other parameters
are ignored.
Configuration
>RS232 settings
Submenu for the serial interface parameters
Open the submenu with
>RS232 settings
Baud Rate:
9600
Transmission speed in baud
300,600,
1200,2400,
4800,9600
>RS232 settings
data bit:
8
Number of data bits
7,8
>RS232 settings
stop bit:
1
Number of stop bits
1,2
>RS232 settings
parity: none
Parity selection
even, odd, none
>RS232 settings
handshake: HWs
Handshake selection
HWs,HWfull,
SWchar,SWline,
none
>RS232 settings
character set:
IBM
Character set for printer or PC
IBM,HP,Epson,
Seiko,Citizen
use
to access the
next highest
level
use
to access the
next highest
level