5 Detailed description
774 Oven Sample Processor, Instructions for use
116
5.11 Description of the remote control commands
5.11.1 &Mode …
5.11.1.1
Mode
$G, $S, $H, $C
Start ($G) or stop ($S) the current method. Interrupt with $H (hold), resume
with $C (continue).
5.11.1.2
Mode.Method read
only
Name of current method in the working memory. $Q sends
8 ASCII characters.
∗∗∗∗∗∗∗∗
stands for a blank method.
5.11.1.3
Mode.SmplNo 1...999,
∗, Rack
Number of samples. This entry determines the number of runs for the sam-
ple sequence.
∗
= infinite number of samples. Processing has to be stopped with
&M;$S or
rack = (Number of rack positions – number of special beakers defined).
5.11.1.4
Mode.StartSeq.1.Cmd
NOP, MOVE, LIFT, SAMPLE,
etc. up to .100
HEATER, DEF, FLOW,
DOS, SCAN, CTRL, WAIT, RACK
Defines the command of the indexed command line in a start sequence.
The introduction of a command appends the according sub-branch from
&Assembly (see page 128ff) to the index node.
A NOP-entry cuts the appended sub-branch from the index node.
Each entry at the end of a sequence appends a new node
&Mode.StartSeq.
∗
.Cmd("NOP").
5.11.1.5
Mode.StartSeq.1.∗
.Move…, .Lift…, .Sample…,
etc. up to .100
.Heater…,.Def…, .Flow…, Dos…,
Scan…, .Ctrl…, .Wait…, .Rack, .Nop
Indexed start sequence; its commands will be executed line by line in proc-
essing. At the index node &Mode.StartSeq.1.Cmd (see 5.11.1.4) the ac-
cording sub-branch from &Assembly… will be appended depending on the
selected command.
Example:
&Mode.StartSeq.1.Cmd("MOVE")
⇒ &Mode.StartSeq.1.Move.Target("1")
⇒ &Mode.StartSeq.1.Move.Position("sample")
5.11.1.6
Mode.SampleSeq.1.Cmd
NOP, MOVE, LIFT, SAMPLE,
etc. up to .100
HEATER, DEF, FLOW,
DOS, SCAN, CTRL, WAIT, RACK
Defines the command of the indexed command line in a sample sequence.
See start sequence (5.11.1.5).