Metrohm 774 Oven Sample Processor User Manual
Page 126
![background image](/manuals/586637/126/background.png)
5 Detailed description
774 Oven Sample Processor, Instructions for use
120
5.11.2.5
Config.RackDef.SpezBeak.1.Pos
0...number of rack positions
etc.
until
.8
Rack positions of special beakers 1 to 8 (position 0 = not defined).
5.11.2.6
Config.PosTab.TabIdx
0...31
Index of position tables..
5.11.2.7
Config.PosTab.Name
8 ASCII characters
Config.PosTab.R1Num
2...(R2Num – 2)
Config.PosTab.R2Num
(R1Num + 2)...(R3Num – 2)
Config.PosTab.R3Num
(R2Num + 2)...200
Config.PosTab.R1Off
0...3599
Config.PosTab.R2Off
0...3599
Config.PosTab.Num
1...200
Definitions of position tables. Depending on the table index the correspond-
ing one of the 16 possible data sets is overlayed.
Name:
Identification of the rack type. Will be available as a selec-
tor under &Config.RackDef.Type (see 5.11.2.4).
R1Num:
Highest Beaker position in row 1
R2Num:
Highest Beaker position in row 2
R3Num:
Highest Beaker position in row 3
R1Off Offset
in
1
/
10
-angular degrees for the beaker positions in
row 1 (for the beaker test)
R2Off Offset
in
1
/
10
-angular degrees for the beaker positions in
row 2 (for the beaker test)
Num:
Number of rack positions
5.11.2.8
Config.PosTab.1.Value
0...3599
bis
.200.Value
Angular offset for the particular rack position in tenth of angular degrees
(
1
/
10
-degrees).
Definition of new rack types:
The definition of rack types is only possible via RS232 interface.
• Set table index (see 5.11.2.6).
• Enter name of rack type (&Config.PosTab.Name, see 5.11.2.7).
• Enter highest beaker position in row 1 to 3 (&Config.PosTab.R1–3Num,
• Enter the offset angle between beaker position 1 in row 1 and beaker
position (R1num+1) in row 2, respectively and tower 1
(&Config.PosTab.R1–2Off, see 5.11.2.7).
• Define number of rack positions (&Config.Pos.Tab.Num, see 5.11.2.7).
• Enter the particular angular offset for each rack position
(Irregular layouts of rack positions are possible)
5.11.2.9
Config.WetPart.WetPartNo
1...12
Identification of dosing unit.
5.11.2.10
Config.WetPart.1.MaxRate 0.01...160 ml/min
Config.WetPart.1.Length 0...1000...30000 mm
Config.WetPart.1.Diameter 0.1...2...20 mm
until
.4.Diameter
(Default value .2.Length : 250 mm)