5.2 The keyboard

774 Oven Sample Processor, Instructions for use

55

Key

Normal state and Editing

0

DEF

SHIFTRATE

Change turning speed and direction
• In addition to giving the turning speed of the sample rack in angular

degrees/sec., the turning direction can also be indicated with the
first parameter.

• Turning direction "+" causes the sample beakers to be processed

in the counterclockwise direction, i.e. in ascending order. Turning
direction "–" indicates in the clockwise direction, i.e. in descending
order. The beaker positions are numbered on every sample rack
and easily visible.

• With turning direction “auto” the sample changer independently

chooses the shortest possible path for placing a vial under a particu-
lar lift. The turning direction is automatically chosen.

• Syntax: SHIFTRATE [Turning direction] [Turning speed]

0

DEF

DRIVE.PORT

Change Dosino port assignments
• The ports for each of the 12 Dosinos which can be connected may

be functionally redefined. Each port can therefore be used as dos-
ing outlet or filling inlet, etc.

• The dosing drive must be given left of the dot of the first parameter

and after the dot, the respective port must be indicated.

• For the second parameter you can choose between the functions:

Dosing (Dos), Filling (Fill), Rinsing (Rinse), Empty (Empty) or Pre-
pare (Prep).

• Syntax: DRIVE.PORT [Dosing drive.Port] [Function]