10 the remote control tree, 1 overview, 2 &mode – Metrohm 774 Oven Sample Processor User Manual
Page 112

5 Detailed description
774 Oven Sample Processor, Instructions for use
106
5.10 The remote control tree
5.10.1 Overview
The internal object tree can be divided into the following branches:
&
Root
├
Mode
Method parameters
├
Config
Instrument configuration
├
Info
Current
Data
├
Setup
Setting the operating mode
├
UserMeth
User-defined methods
├
Assembly
Component data
└
Diagnose
Diagnostics program
Detailed Description of the Main Branches:
5.10.2 &Mode
Object Description
Input
range
Reference
& Root
├
Mode
Method parameters
$G, $S, $H, $C
:
├
.Method
Method name
8 ASCII characters
├
.SmplNo
Number of samples in a series
1...999, ∗, Rack 5.11.1.3
│
├
.StartSeq
Start sequence -
│ ├
.1
Line number of the command
-
│ │
├
.Cmd Command
nop, MOVE, LIFT,
│ │
└
.
∗
SAMPLE, HEATER, DEF
│ │
FLOW, DOS, SCAN,
│
∗
CTRL, WAIT, RACK
│ :
│ └
.100 Sequence
end
NOP
│
│
├
.SampleSeq
Sample sequence -
│ ├
.1
Line number of the command
-
│ │
├
.Cmd Command
nop, MOVE, LIFT,
│ │
└
.
∗
SAMPLE, HEATER, DEF
│ │
FLOW, DOS, SCAN,
│
∗
CTRL, WAIT, RACK
│ :
│ └
.100 Sequence
end
NOP
│
: