3.3 Methods and Sequences
774 Oven Sample Processor, Instructions for use
35
The LEARN function is available for the following commands:
Command Adaptive parameter
Type of function
LIFT
Lift position in mm
absolute
WAIT
pause time in seconds
additive
DOS
Dosing volume in mL
additive
SCN Rm
Status of the 8 remote lines
"live" value
SCN RS
Received character string
"live" value
TRACE function
The "TRACE" function is a valuable aid for processing a complete sequence
or method, or excerpts, step-by-step for test purposes. Each command line
in a sequence can be carried out directly by pressing the
When the action is finished the following command line is displayed.
"Tracing" can be used directly after the entry of a command line or at any
time after the parameter menu has been opened and a sequence has been
selected.
3.3.3 Process control
rupted manually and if no unexpected errors occur the sample series will be
processed properly and terminated with the final sequence. The sample
sequence is carried out several times according to the entry in 'No. of sam-
ples', starting with the sample beaker which has been defined as 'SAM-
PLE'.
If the sample series is interrupted with
sor immediately returns to the normal state. Unprocessed samples are not
taken into account; the final sequence is not carried out. If settings for this
occurrence have been entered under '>manual stop' then the correspond-
ing actions or commands will be carried out via the interface in order to
stop the connected instruments as well or to carry out other actions.
The sample series can also be interrupted with
which is currently active is immediately terminated.
to continue the method with the following command in the sequence which
is active. Connected peripheral units are not stopped with the
key. In the 'HOLD' condition the current oven temperature is maintained.
has ended (soft termination). The sample which is currently being proc-
essed will be processed to the end.
the next command line of the sequence.
If errors occur during a sample series then a corresponding error mes-
sage will appear; this must be confirmed with
changer then goes to the HOLD status (see above). When the error has
been remedied
If the gas flow limits are exceeded or undercut this does not interrupt the
method sequence. The gas flow can be corrected manually with the gas
flow regulator.